Abstract: This article proposes a novel cuff-based microactuator for automatic blood extraction using
Shape Memory Alloy (SMA) technology. It aims to provide an actuator solution for an electronic
mosquito-like device, e-Mosquito, which enhances blood glucose measurement to off the potential to
significantly improve both the quality of life for diabetic patients and their disease management.
Fabrication and assembly methods of the microactuator prototype are discussed. The 6-mm thick
microactuator prototype underwent mechanical tests in a laboratory environment to characterize its
penetration force and depth, which are the major factors facilitating skin penetration and blood
extraction. Testing results demonstrated that this actuator can produce a maximum penetration force of
160gf and a maximum displacement depth of 2.45mm, which both exceed the minimum force and depth
requirements for blood capillary-reaching skin penetration in humans. Five actuator prototypes were
assembled and integrated to a cuff wrist strap. A pilot human study was performed to test the blood
extraction capabilities of the device. All five actuators successfully penetrated the skin and drew whole
blood samples appearing on the skin surface of the subject.